KANGAROO JOEY
11
Enteral Feeding Pump
Figure 7A. Opening menu
for Continuous Feed mode
with a Feed-Only set.
Figure 7B. Opening menu
for Continuous Feed mode
with a Feed & Flush set.
Figure 7C. Opening menu
for Intermittent Feed mode with a
Feed-Only set.
Figure 7D. Opening menu
for Intermittent Feed mode
with a Feed & Flush set.
Prime Pump
The KANGAROO JOEY Enteral Feeding Pump may be primed automatically with a single button push, including the flushing line if a
Feed & Flush Set is loaded. The pump can also be primed in a more interactive method using the hold-to-prime feature.
After a pump set has been loaded and the status line shows SET LOADED, press4PRIME PUMP in the opening menu (Figures 7A-
7D) to get one of the PRIME PUMP menus shown in Figures 8A and 8B.
Figure 8A. PRIME PUMP
menu for Feed-Only set.
Figure 8B. PRIME PUMP
menu for Feed & Flush set.
Auto Priming
The AUTO PRIME option will be available if the pump senses that the currently installed pump set was not previously primed, and
that there is no fluid in the line near the rotor. If auto-priming is desired, press and release4AUTO PRIME to automatically prime
the line(s). For Feed & Flush Pump Sets, the auto prime feature will automatically prime both lines, starting with the flush line.
Press4STOP to cancel AUTO PRIME.