KANGAROO JOEY
™
12
Enteral Feeding Pump
When auto-priming is completed, the status line will show AUTO PRIME COMPLETE and will no longer show the AUTO PRIME
option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the priming, as described below.
Hold-to-Prime
The hold-to-prime options allow for precise interactive control of the priming process.
Feed-Only Pump Sets
For Feed-Only Pump Sets, the screen will appear as in Figure 8A. Press and hold4HOLD TO PRIME FEED until the feed line has
been primed down to the stepped end connector at the end of the pump set.
Feed & Flush Pump Sets
For Feed & Flush Pump Sets, the display will appear as in Figure 8B. First press and hold4HOLD TO PRIME FLUSH to prime the
flushing line, and then press and hold4HOLD TO PRIME FEED to prime the feed line down to the stepped end connector at the end
of the pump set.
If a Feed & Flush Pump Set is used, remember to first press4HOLD TO PRIME FLUSH until the flushing fluid has reached the valve,
and then press4HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the stepped end connector.
If the feeding line is primed first, subsequent priming of the flushing line will force the air between the flush solution bag and valve
into the main line, which already has been primed with enteral formula.
Re-priming after Bag Empty
A pump set bag that has been emptied will trigger the FEED ERROR screen. In this condition the pump set bag can be refilled to
continue the feeding, but only after the pump set has been re-primed.
To re-prime the pump set, do the following:
• Disconnect the feeding line from the patient
• Refill the bag
• Press4CONTINUE to begin the pump RUNNING
• Press4HOLD
• Press4ADJUST SETTINGS
• Press4PRIME PUMP
• Press4HOLD TO PRIME to prime the line(s)
Pump sets should not be reused after 24 hours of initial usage.
Selecting Feed Mode
EZ Pump Mode Option
The EZ PUMP MODE option is a limited, CONTINUOUS MODE only setting of KANGAROO JOEY Enteral Feeding Pump. It feeds
non-stop (no VTBD option) at a rate 0-400mL/hr as defined by the user until manually stopped. It has no BOLUS, FLUSHING, RUN
MODE SCREEN LOCK, or AUTORESUME capability and displays no HISTORY information.