KANGAROO JOEY
17
Enteral Feeding Pump
Figure 14A. ADJUST FLUSH
menu, intermittent mode.
Figure 14B. Setting the Flush
Volume.
Figure 14C. Setting the Flush Time
Interval.
Note: The nature of two intermittent processes, periodic bolus feeding and periodic flushing, is such that occasional conflicts can
arise where each process needs to be running at the same time. The flushing process is assigned precedence over the feeding process
for Feed & Flush pump sets. “Flushing precedence” means that a flush cycle will run to its end even if a bolus delivery is scheduled
to begin. It also means that a flush cycle will interrupt a bolus feeding cycle that is already in progress. The feeding cycle will pick
up where it was interrupted after the flushing cycle is complete. It is therefore possible that the net feeding volume is reduced from
what may be intended, over a given time period.
Press4FLUSH VOLUME in Figure 14A to define the volume per flush cycle. Press the4buttons in Figure 14B to program the flush
volume from 10 to 500 in increments of 1 mL. Press4ENTER (top button) to exit the menu.
Press4FLUSH INTERVAL in Figure 14A to define the time interval between starts of flushing. Press the4buttons in Figure 14C to
program the time interval from 1 to 24 hours in increments of 1 hour. Press4ENTER to exit the menu.
The pump will automatically limit flushing settings which exceed pump capabilities to deliver.
CAUTION: The KANGAROO JOEY Enteral Feeding Pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when
programming the flush volume so that it matches the patient’s need.
NOTE: The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as
long as at least one flush had occurred during formula delivery.
Run
The KANGAROO JOEY Enteral Feeding Pump will only run when a pump set is properly loaded and the necessary feeding parameters
have been programmed. For a Feed & Flush Set, the flushing parameters do not have to be set for the pump to begin operating. If
flushing parameters are set as zeros, there will be no flushing.
After the pump has been programmed, press4RUN to start operation. Figure 15A shows the RUNNING screen for continuous
feeding with flushing. The status line will indicate RUNNING, and the droplet icon will move vertically on the screen. The green LED
indicator will be illuminated to give a quick visual indication of positive status in a darkened room.
The RUNNING screen shows the feed rate, amount fed, flushing data if a Feed & Flush Pump Set is used, and the remaining volume
to be delivered (remaining VTBD). The remaining VTBD is shown next to the
icon.
The RUN option can be selected from the following screens, if the proper parameters are programmed and the pump set is loaded:
SET LOADED screens (Figures 7A-7D)
ADJUST FEED screen, continuous mode (Figure 11A)
ADJUST FLUSH screens (Figure 12A, 14A)
HOLDING screens (Figure 18A)