KANGAROO JOEY
™
20
Enteral Feeding Pump
Flush Now (Not Available in EZ Pump Mode)
During the RUNNING mode, the pump can be diverted to immediately perform a flush (flush on demand) when a feed/flush set
is loaded by pressing4FLUSH NOW in the RUNNING menu, input a desired flush volume, then press4ENTER. See Figure 15A.
When4FLUSH NOW is pressed, the screen will prompt for the volume to be immediately flushed, and will show the previously
programmed flush volume as default but may be adjusted to the users desired volume. The screen will appear as the standard
Flush input screen as shown in Figure 12B (or 14B). Setting the FLUSH NOW flushing volume is temporary and will not change the
main/periodic flushing volume previously programmed. If4FLUSH NOW is pressed by accident; the user may either press the cancel
button or allow the screen to return to the RUNNING screen after 10 seconds, if there is no input for the flush volume.
If stopping a FLUSH NOW before the set volume has been delivered is desired, press4FLUSH NOW and enter a volume of zero and
press4ENTER. The FLUSH NOW will stop immediately and the pump will return to the RUNNING mode.
More Options
Press4MORE in one of the opening menus, Figures 5A-5B or Figures 7A-7D, to access the MORE OPTIONS screen, Figure 19.
Figure 19. MORE
OPTIONS menu.
Buzzer
Press4BUZZER in the MORE OPTIONS menu to access the BUZZER screen and increase or decrease the alarm buzzer volume. The
alarm will be audible as the volume setting is changed. This sets the volume level for any warning or error that may occur with the
pump. Critical errors of System Error and Pump Set Dislodged will always use the maximum volume setting.
History (Not Available in EZ Pump Mode)
Press
4HISTORY in the MORE OPTIONS menu to access the HISTORY screen. Press the4buttons in Figure 20 to set the time history
of interest. The totals of FEED mL and FLUSH mL will be shown for the requested history time.
WARNING: A maximum of 72 hours of past history can be retrieved, excluding data from the current hour. NOTE: The 72-hour
history includes time when the pump is powered OFF.