KANGAROO JOEY
™
 
 
29
Enteral Feeding Pump
Figure 28.  Feed Error screen.
Flush Error (Adjustable Audible Volume)
The FLUSH ERROR screen appears when the flushing solution is no longer being delivered because the flush bag is empty or 
because of a clog between the pump and flush bag.  Check the flush line to find the occlusion causing the blockage.  If the occlusion 
cannot be cleared, load a new pump set, prime it, and press
4CONTINUE to restart the feeding, Figure 29.
If the error still cannot be resolved, press    POWER to stop operation of the pump and put a different pump into service.
Figure 29.  Flush Error screen.
Flow Error  (Adjustable Audible Volume)
The FLOW ERROR screen appears when either the feeding or the flushing solution is no longer being delivered because of a clog 
between the pump and the patient.  Check the line to find the occlusion causing the blockage.  If the error cannot be resolved, check 
the valve pocket in the pump set loading area for dampness or dirt, which may be a cause for a false error.  Clean and dry the valve 
pocket.  If the error still cannot be resolved, load a new pump set, prime it, and press4CONTINUE to restart the feeding, Figure 30.
If the error still cannot be resolved, press    POWER to stop operation of the pump and put a different pump into service.
Figure 30.  Flow Error screen.
Pump Set Dislodged Error  (Fixed Audible Volume)
The PUMP SET DISLODGED screen will appear if the black ring retainer (MISTIC) is not properly loaded in the MISTIC pocket in the 
pump set loading area.  Check and correct the positioning of the MISTIC retainer if possible.  If the error cannot be resolved, load a 
new pump set, prime it, and press4CONTINUE to restart the feeding, Figure 31.
If the error cannot be resolved, press    POWER to stop operation of the pump and put a different pump into service.