KANGAROO JOEY
4
Enteral Feeding Pump
of this equipment, may lead to a reduced level of safety of the pump. While use of the pumps serial port is not recommended for
customers, the following should be considered in the event that the customer decides to connect to this port:
• The accessory is specifically certified as complying with the safety requirements for accessories used with the pump, whereas the safety
certification of the accessory has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized standard.
• Under no circumstances should the certified accessory be connected to the serial I/O port while pump is in use for enteral feeding of a
patient.
14. CAUTION: This enteral feeding pump should only be used for patients who can tolerate the flow rates and accuracy levels delivered by the
pump. Premature infants may require higher accuracy rates than specified for this enteral feeding pump. For optimal accuracy, the top
of the starting volume of formula should be 6 inches above the pump, do not reuse feeding sets and avoid overstretching
the silicone that wraps around the pump rotor.
15. Do not attempt to rotate the valve stem inside the feeding set valve assembly. Rotation of the valve stem can only be done by the pump
while the valve assembly is properly loaded into the pump.
16. Should feeding sets require rinsing, it is recommended that the feeding sets be rinsed while they are loaded in the pump.
17. The KANGAROO JOEY Enteral Feeding Pump has been programmed to optimize accuracy by taking into account the viscosity
of formula. Formula that has been diluted and has a low viscosity will tend to deliver like water and thus deliver to the
high side of the accuracy specification.
18. CAUTION: Ensure buzzer hole is unobstructed during normal operation so as to allow clear recognition of alarm.
19. CAUTION: The battery cells used in this device may present a fire or chemical hazard if mistreated. Do not disassemble, heat above 100°C
(212°F), or incinerate.
Section III – Icon Identification
Run Mode
Vertically-moving droplet in
RUNNING screen indicates Run
mode.
Battery Symbols
• Full charge
• 2/3 or less charge
• 1/3 or less charge
Very low battery (15 minutes max.)
Auto Priming Indicator
When auto priming, the bar will progress
to show auto prime feature is active.
Volume Setting
Indicates alarm volume setting.
Error Indicator
Pump operations are stopped
until the error is resolved.