KANGAROO JOEY
14
Enteral Feeding Pump
Select4CONTINUOUS/INTERMITTENT under the MORE OPTIONS menu, Figure 10A. Then press 4CONTINUOUS MODE
or4INTERMITTENT MODE from the SELECT MODE menu, Figure 10B. Press4DONE to exit.
Figure 10A. MORE OPTIONS menu.
Figure 10B. SELECT MODE menu for
continuous or intermittent mode
selection.
Continuous Mode Adjust Feed
Continuous Mode Feeding
For CONTINUOUS feeding mode, the ADJUST FEED option is available in the opening menus (Figure 7A for Feed-Only sets or Figure
7B for Feed & Flush sets). Select this option to program the feeding parameters from the ADJUST FEED screen, Figure 11A.
Press4FEED RATE in Figure 11A to define the rate of delivery. Press the4buttons in Figure 11B to program the rate from 1 to 400
in increments of 1 mL/hr. Press4ENTER (top button) to exit the menu.
Press4FEED VTBD in Figure 11A to define the Volume To Be Delivered. Press the4buttons in Figure 11C to program the volume
from 1 to 3000 mL in increments of 1 mL. Press4ENTER (top button) to exit the menu. If the FEED VTBD is set to zero, the pump
will run until the supply is exhausted.
NOTE: When a FEED VTBD is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to the
icon on
the lower right-hand portion of the screen (Figure 11A). If the FEED VTBD is reprogrammed during a feeding, from HOLDING mode,
the pump will restart the feeding and deliver the entire volume of the new VTBD setting.
Figure 11A. ADJUST FEED menu.
Figure 11B. Setting the Feed Rate. Figure 11C. Setting the Feed
Volume to be Delivered.
Continuous Mode Flushing (Not Available in EZMODE)
For CONTINUOUS feeding mode with a Feed & Flush Set loaded, the opening menu will appear as shown in Figure 7B.
Press4ADJUST FLUSH in Figure 7B to program the flushing parameters from the ADJUST FLUSH screen, shown in Figure 12A.
Press4FLUSH VOLUME in Figure 12A to define the volume per flush occurrence. Press the4buttons in Figure 12B to program the
flush volume from 10 to 500 in increments of 1 mL. Press4ENTER (top button) to exit the menu.