Section VII — Cleaning 
Cleaning should be performed as needed.  It may also be desirable to define cleaning intervals based on knowledge of the 
environment in which the pump is used.  Only personnel trained in the cleaning of medical devices should perform cleaning.
CAUTION: Do not immerse pump or power cord in water or other cleaning solution; clean using a damp cloth or 
sponge. Failure to follow the cleaning procedures described herein could result in hazards to users.  As with any AC 
powered electrical device, care must be taken to prevent liquid from entering the pump to avoid electrical shock 
hazard, fire hazard, or damage to electrical components.
If any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel trained in 
servicing KANGAROO JOEY Enteral Feeding Pumps:
•  wetting of the pump’s power cord or leakage into the pump interior during cleaning.
General Cleaning Directions
Cleaning of KANGAROO JOEY Enteral Feeding Pumps must be performed as follows:
CAUTION: Disconnect pump from AC power source before cleaning.  After cleaning, do not connect to AC power source 
until pump and power cord are thoroughly dry.
•  A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and hypochlorite 
mixture, however, repeated cleaning with this solution can damage the plastic pump case. Isopropyl alcohol, applied with a damp 
cotton swab, may be used for cleaning difficult-to-reach areas, however, it should be used sparingly because repeated cleaning 
may damage the case. Allow excess moisture to evaporate from the cord prior to use of power cord.
•  DO NOT USE strong cleaners such as SPRAYNINE™*, PHISOHEX™*, HIBICLENS™*, or VESTASYDE™* because damage to the pump 
case housing can result.
Directions for Cleaning Pump Housing
•  Refer to General Cleaning Directions before starting.
•  Clean outside surface with a damp cloth or sponge.
Directions for Cleaning Pump Power Cord
•  Refer to General Cleaning Directions before starting.
•  Unless soiling is observed, the power cord should not be cleaned.
•  If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with  a cloth 
dampened with isopropyl alcohol.
CAUTION: Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or fire hazard.