KANGAROO JOEY
™
19
Enteral Feeding Pump
Hold
Pump operation can be paused by pressing4HOLD while it is RUNNING, such as in Figure 15A. The status line will indicate
HOLDING and the yellow LED indicator will be lit up to give a quick visual indication of an informational status in a darkened room.
Note: the pump’s alarm will sound if it has been left HOLDING for 10 minutes without further input.
There are five options available during HOLDING. See Figures 18A and 18B.
Figure 18A. HOLDING mode
options, continuous feeding mode.
Figure 18B. HOLDING mode
options, intermittent feeding mode.
Clear Volume Fed
Press4CLEAR VOLUME to re-zero the output feed totals “mL FED,” “BOLUSES FED,” and “mL FLUSH.”
Adjust Settings
Press
4ADJUST SETTINGS to return to one of the opening menus shown in Figures 7A-7D, thus allowing any setting to be modified.
If the pump set has been removed after going into HOLD, the opening menu will show LOAD A SET, as seen in Figures 5A-5B
.
Resume In __ Minutes (Not Available in EZ Pump Mode)
Press4“RESUME IN __ MIN” once to immediately place the KANGAROO JOEY Enteral Feeding Pump into a 30-minute resume
running mode. Pressing the4RESUME IN __ MIN button again will cyle the “resume in” time back to zero. No other input is needed
to start the countdown, as the time immediately counts down after the 30-minute value is reached.
Run
Press4RUN to immediately restart pump operation. This option will not be available if the pump set is removed during HOLD.
History (Not Available in EZ Pump Mode)
Press4HISTORY to access the HISTORY screen. Use this screen to review feeding and flushing history. See subsection “More
Options” below for further information.