KANGAROO JOEY
™
 
 
16
Enteral Feeding Pump
NOTE:  If the NUMBER OF BOLUSES is reprogrammed during a feeding (from HOLD mode), the pump will restart the bolus feeding 
and will deliver all the boluses as defined by the new setting.
Figure 13A.  ADJUST BOLUS 
menu.
Figure 13B.  Setting the Bolus Rate.
Figure 13C.  Setting the Number of 
Boluses.
Figure 13D.  Setting the volume per 
Bolus.
Figure 13E.  Setting the time interval 
between starts of bolus.
Intermittent Mode Feeding – Bolus Max  (Not Available in EZ Pump Mode)
The BOLUS MAX option can be used to deliver a bolus at a high rate, similar to gravity feeding.  
Press4BOLUS MAX in the BOLUS RATE menu, Figure 13B, to program the BOLUS MAX mode.  
The feeding rate for BOLUS MAX is 999 mL/hr.  
Intermittent Mode Flushing 
For INTERMITTENT feeding mode with a Feed & Flush Set, the opening menu, shown in Figure 7D, will show the ADJUST FLUSH 
option.  Press4ADJUST FLUSH in Figure 7D to program the flushing parameters from the ADJUST FLUSH screen, Figure 14A.