UTD2000 Series User Manual 
31 
 
Chapter V   Acquire System 
  User can change the data acquisition setting of the DSO by pressing the ACQUIRE 
button as shown in the figure below UTD2000 series DSO.   
 
Figure 5-1 Function Key of Acquire System 
While ACQUIRE button is pressed, the function menu will change to set the data 
acquisition setting for the DSO. 
Table 5-1 Acquire Function Menu 
Function Menu  Setting  Description 
Acquisition  
Mode 
Sampling    Set to sampling acquisition mode.   
Peak value  Set to peak value acquisition mode.   
Average    Set to average acquisition mode.   
Averages   2~256 
At the average sampling mode, set average number by using the multipurpose 
knob (Average number can set to 2,4,8,16,32,64,128,256)   
Sampling 
Real-time    Set to Real-time sampling.   
Equivalence    Set to Equivalent sampling.   
Fast Acq 
ON  Turn on fast acquiring. Waveform capturing rage up to at 2k waveform/second. 
OFF  Turn off fast acquiring. 
āā   
By changing the acquisition setting of the DSO, user can observe the input signal at 
different manner.    Example below is a noise waveform acquired by different acquisition 
setting of the UTD2000 series DSO.     
Figure 5-1 shows the original signal using sampling acquisition, user can see that the 
waveform display is very noise.     
Figure 5-2 shows the same signal with average acquisition of a 32 number, please 
observe that the waveform display become much more smooth.   
   
Figure 5-1 using sampling method      Figure 5-2 using average acquisition of 32 number