UTD2000 Series User Manual 
50 
 
example, if the pulse is a logical signal of TTL level, the trigger level should be set at about 
2V and the trigger edge should be set to rising edge trigger. If you
 are not certain about 
the signal, you can observe it by automatic or
 normal trigger to determine the trigger level 
and trigger edge. 
Steps : 
(1) As in the previous illustration, set the attenuation factor of the
 probe and CH1. 
(2) Carry out trigger setup 
①  Press TRIG MENU in the trigger control zone to display the trigger setup 
menu. 
②  In this menu, use F1~F5 set the trigger type to EDGE, set trigger source to 
CH1, set inclination to Rising, set trigger type to Single and set trigger coupling to AC. 
③  Adjust horizontal time base and vertical range to an appropriate range. 
④  Turn the TRIGGER LEVEL
 control knob to get the desired 
⑤  Press RUN/STOP and wait for a signal that meets the trigger
 condition. If any 
signal reaches the set trigger level, the system will sample once and display it on the 
screen. By
 using this function you can easily acquire any occasional event. For 
example, when a sudden glitch of relatively big
 amplitude is acquired : set the trigger 
level to just higher than
 the normal signal level. Press RUN/STOP and begin waiting. 
When a glitch occurs, the machine will automatically trigger and record the waveform 
immediately before and
 after triggering. By turning the horizontal POSITION knob in 
the horizontal control zone on the front panel, you can change the trigger position 
horizontally to achieve negative delay
 trigger of various lengths for easy observation 
of waveform
 occurring before the glitch 
 
Figure 12-3 Single signal 
 
Example 4: Reducing random noise of signals 
If the signal being measured is stacked with random noise, you can adjust the setups 
of your oscilloscope to filter or reduce the noise, so it will not cause interference to the 
signal during measurement. (Waveform is shown below)