Programable DC Electronic Load User Manual
73
Remarks:
Environmental requirements: 23℃±5℃ ≤90%RH.
Recommended calibration frequency: once/year
AC power input level: (The switch on the rear panel of the electronic load can choose
110V or 220V)
110V: 110V±10% 50~60Hz
220V: 220V±10% 50~60Hz
Current
(CC)
≤3A ≤30A ≤3A ≤30A ≤3A ≤15A
35mΩ 35mΩ 35mΩ 35mΩ 35mΩ 35mΩ
≥1.1 times the set value,
immediate protection
Over temperature
protection
Impedance of input
terminal
Requirements of power
supply
Size
mm(Width*Height*Depth)
≥1.05 times the set value, delay
protection
≥1.1 times the set value,
immediate protection
>(FS+15W), immediate protection
≥1.05 times the set value, delay
protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value,
immediate protection
>(FS+15W), immediate
protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value,
immediate protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value, immediate
protection
>(FS+3V), immediate protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value, immediate
protection
>(FS+15W), immediate protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value, immediate
protection
>(FS+1.5A), immediate protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value,
immediate protection
>(FS+3V), immediate protection
>(FS+1.5A), immediate
protection
>(FS+1.5A), immediate protection
≥1.05 times the set value, delay
protection
≥1.1 times the set value,
immediate protection
>(FS+3V), immediate protection