Chapter 4: Operating Instructions
UT100 Handheld Pulse Oximeter Operation Manual 4-3
while the monitor is turned off.
4.4 Attaching the Sensor to the Patient
WARNING! Prolonged use or the patient’s condition may require
changing the sensor site periodically. Change sensor site
and check skin integrity, circulatory status, and correct
alignment at least every 4 hours.
Attaching the patient to the monitor requires these
st
eps:
1. Choose the
sensor
.
2. Check the sensor and oximetry
cable
.
3. Clean or disinfect the sensor if
using the reusable type (Disposable sensors
are for
single
-
patient use and do not require cleaning or
disinf
ec
ting).
4. Attach the sensor to the
pa
tien
t.
4.4.1 Choosing the Sensor
WARNING! Before use, carefully read the sensor directions for use,
including all warnings, cautions, and instructions.
Choose the
appropriate
sensor from the following
char
t.
PATIENT
SITE
DESCRIPTION
Adult
45Kg
Finger
Finger or toe
Ear
Sensor, Adult (reusable)
Sensor, Disposable, Adult Finger
Sensor, Ear (reusable)
Pediatric
15-45Kg
Finger
Finger or toe
Ear
Sensor, Adult (reusable)
Sensor, Disposable, Ped. Finger
Sensor, Ear (reusable)
Infant
3-15Kg
(for spot-check only)
Hand or Foot
Toe
Finger or Toe
Sensor, Universal
Y
(reusable)
Sensor, Wrap, Infant (reusable)
Sensor, Disposable, Infant
Neonate
3Kg
(for spot
-
check only)
Hand or Foot
Foot
Sensor, Disposable, Neonate
Sensor, Wrap. Neonate(reusable)