Chapter 8: Maintenance and Troubleshooting
UT100 Handheld Pulse Oximeter Operation Manual 8-1
Chapter 8: Maintenance and Troubleshooting
8.1 Schedule of Maintenance
MAINTAIN THIS ITEM
HOW OFTEN
BY
D
O
I
N
G
T
H
I
S
Ba
tt
er
y
When
Ba
tt
er
y Level icon
is
flashing
,
and/or
audible
alarm
sounds
.
Follow the instructions for
installing the
ba
tt
er
ies
.
Disinfecting the
r
eusable
sensor
.
B
ef
or
e
attaching
the
sensor to the
pa
tien
t.
Follow the instructions for
cleaning the reusable sensor.
Disinfecting the
monit
or
.
When
nec
essar
y
.
1.
R
emo
v
e
the batteries from the
unit.
2. Wipe the surfaces of the
monitor with a
soft,
clean cloth
dampened
in isopropyl
alc
ohol
.
Use only a cloth that is
dampened
,
not
w
et.
CAUTION
!
Do not allow isopropyl alcohol or water to enter any of the
openings on the monitor. Evidence that liquid has been
allowed to enter the monitor voids the warranty.
8.2 Storage
WARNING! To ensure accurate performance and prevent device failure,
do not subject the UT100 to extreme moisture such as
direct exposure to rain. Such exposure may cause
inaccurate performance or device failure.
Whenever
possible
,
the monitor should be stored at room
temperature
in a
dry
en
vir
onmen
t.
If it is necessary to store the monitor for an
extended
period of
time
,
the
unit should be
packed
in
its original shipping
c
on
tainer
.
Storing the monitor
for a long period of time may
degrade
the
battery
capacit
y
.
Batteries should
be removed from the monitor before
st
oring
.
Storage specifications are as follows: