Chapter 6: Trend Data Output and Analysis
UT100 Handheld Pulse Oximeter Operation Manual 6-7
Low SpO
2
: The lowest SpO
2
value taken from the valid sample with the pulse
rate and time at which the lowest SpO
2
value occurred.
SpO
2
Standard Deviation: The measure of the variability of the sample. A
higher standard deviation indicates greater variability in the samples.
High PR: The highest pulse rate value taken from the valid sample and the
SpO
2
value& time the highest pulse rate occurred.
Low PR: The lowest pulse rate taken from the valid sample and the SpO
2
value
& time the lowest pulse rate occurred.
2. Desaturation Report: Lists up to 50 of the deepest (lowest SpO
2
values)
desaturation events. The events are listed in order of depth, beginning with
the deepest event.
Figure 6.7: The Desaturation Report
Reference for the event: A unique number between 1-50, which identifies
each desaturation event.
Duration of the event: calculated by the software.