Vaisala HMP4 - Figure 39 HMP4 Humidity Measurement Accuracy as a Function of Temperature; Figure 40 HMP4 Temperature Measurement Accuracy over Full Range

Vaisala HMP4
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Property Description/Value
Temperature
Measurement range −70 … +180 °C (−94 … +356 °F)
Accuracy
1)
4)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
2)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Sensor Pt100 RTD Class F0.1 IEC 60751
1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) Chemical purge feature available with this sensor
4) Exposing temperature sensor to temperatures outside −20 … +150 °C (−4 … +302 °F) may
cause permanent additional deviation of ±0.1 °C (0.18 °F)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Accuracy [+/- % RH]
Temperature [
°
C]
3.5
130
140
150
160
170
180
4.0
90 ... 100 %RH
0 ... 90 %RH
Figure 39 HMP4 humidity measurement accuracy as a function of temperature
0
0.1
0.2
0.3
0.4
Accuracy [+/-
°C]
Temperature [
°
C]
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
0.5
-70
-60
-50
Figure 40 HMP4 temperature measurement accuracy over full range
Chapter 9 – Technical data
71