EasyManua.ls Logo

Vaisala HMP4 - Figure 42 HMP5 Humidity Measurement Accuracy as a Function of Temperature; Figure 43 HMP5 Temperature Measurement Accuracy over Full Range; Table 25 HMP5 Operating Environment

Vaisala HMP4
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Accuracy [+/- % RH]
Temperature [
°
C]
3.5
130
140
150
160
170
180
4.0
90 ... 100 %RH
0 ... 90 %RH
Figure 42 HMP5 humidity measurement accuracy as a function of temperature
0
0.1
0.2
0.3
0.4
Accuracy [+/-
°C]
Temperature [
°
C]
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
150
160
170
180
0.5
-70
-60
-50
Figure 43 HMP5 temperature measurement accuracy over full range
Table 25 HMP5 operating environment
Property Description/Value
Operating temperature of probe body −40 … +80 °C (−40 … +176 °F)
Operating temperature of probe head −70 … +180 °C (−94 … +356 °F)
Operating environment Suitable for outdoor use
IP rating of probe body IP66: Dust-tight. Protected from powerful water
jets from any direction.