EasyManua.ls Logo

Vaisala HMP4 - Table 37 HMP8 Operating Environment; Table 38 HMP8 Inputs and Outputs; Table 39 HMP8 Output Parameters

Vaisala HMP4
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 37 HMP8 operating environment
Property Description/Value
Operating temperature of probe body −40 … +80 °C (−40 … +176 °F)
Operating temperature of probe head −70 … +180 °C (−94 … +356 °F)
Operational pressure < 40 bar
Operating environment Suitable for outdoor use
Measurement environment For air, nitrogen, hydrogen, argon, helium,
oxygen, and vacuum
1)
IP rating of probe body IP66: Dust-tight. Protected from powerful water
jets from any direction.
1) Consult Vaisala if other chemicals are present. Consider safety regulations with flammable
gases.
Table 38 HMP8 inputs and outputs
Property Description/Value
Operating voltage 15 … 30 V DC
Current consumption 10 mA typical, 500 mA max.
Digital output RS-485, non-isolated
Protocols Modbus RTU
Table 39 HMP8 output parameters
Output parameter name and unit
Absolute humidity (g/m
3
)
Relative humidity (%RH)
Absolute humidity at NTP (g/m
3
)
Relative humidity (dew/frost) (%RH)
Dew point temperature (°C) Temperature (°C)
Dew/frost point temperature (°C) Water concentration (ppm
v
)
Dew/frost point temperature at 1 atm (°C) Water concentration (wet basis) (vol%)
Dew point temperature at 1 atm (°C) Water mass fraction (ppm
w
)
Dew point temperature dierence (°C) Water vapor pressure (hPa)
Enthalpy (kJ/kg) Water vapor saturation pressure (hPa)
Mixing ratio (g/kg) Wet-bulb temperature (°C)
Chapter 9 – Technical data
83

Table of Contents

Other manuals for Vaisala HMP4

Related product manuals