Vaisala HMP4 - Figure 45 HMP7 Humidity Measurement Accuracy as Function of Temperature

Vaisala HMP4
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Property Description/Value
Sensor options HUMICAPâ R2
HUMICAPâ R2C
3)
HUMICAPâ 180VC
3)
4)
Temperature
Measurement range −70 … +180 °C (−94 … +356 °F)
Accuracy at +23 °C (+73.4 °F)
1)
5)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
2)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Sensor Pt100 RTD Class F0.1 IEC 60751
1) Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) Chemical purge feature available with this sensor
4) H
2
O
2
resistant. With HUMICAP
â
180VC sensor, accuracy is not specified below −20 °C
(−4 °F) operating temperature.
5) Exposing temperature sensor to temperatures outside −20 … +150 °C (−4 … +302 °F) may
cause permanent additional deviation of ±0.1 °C (0.18 °F)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
-40
-30
-20
-10
0
10
20
30
40
50
60
70
80
90
100
110
120
Accuracy [+/- % RH]
Temperature [
°
C]
3.5
130
140
150
160
170
180
4.0
90 ... 100 %RH
0 ... 90 %RH
Figure 45 HMP7 humidity measurement accuracy as function of temperature
HMP Series with MMP8 and TMP1 User Guide M212022EN-F
78