Chapter 8 ______________________________________________________________ Maintenance 
 
VAISALA ______________________________________________________________________ 105 
CHAPTER 8 
MAINTENANCE 
This chapter contains information that is needed for performing visual 
inspection, cleaning, and verifying the operation of WMT700. 
WARNING 
Some WMT700 product versions provide heating for transducers and/or 
array arms. To avoid injury, do not touch the heated parts of the wind 
sensor when the heating is enabled.  
 
CAUTION 
When handling WMT700, do not rotate, pull, strike, bend, scrape or 
touch the transducers with sharp objects. Any impact on the wind sensor 
array damages the device. 
 
Periodic Maintenance 
WMT700 is a very reliable and rugged sensor. Since no moving or 
consumable parts are used, no periodic maintenance is required. 
WMT700 has been calibrated at the factory, and you are not required to 
perform recalibration.  
NOTE 
Periodic calibration of WMT700 is not required for technical reasons. 
You can use the WM Verifier to test the mechanical integrity of the 
device and perform one-point calibration. 
However, some quality management systems may require regular 
calibration of the measuring instruments.  
To fulfill these requirements, Vaisala recommends that you recalibrate 
the wind sensor every 24 months. Contact Vaisala Service Center for 
more information.