47
11.4. Aiming beam not correctly centered
Press reset on the treatment information window
If the aiming beam is still not centered contact Valon
Lasers or local distributor’s service for help.
11.5. Smart wheel not active
If smart wheel is not, active try restarting the system.
The smart wheel needs to be connected to the system
be-fore the system starts. Otherwise it doesn’t work
If the smart wheel does not work after restarting the sys-
tem, contact Valon Lasers or local distributor’s service
for help.
11.6. Foot switch not active
Set the system to ”Laser disabled”- mode and enable
the laser again.
Check that the foot switch cable is connected to the
laser console
Restart the system.
If the foot switch still doesn’t work, contact Valon Lasers
Service.
11.7. Error messages
In case the program displays an error message:
See section 10 for more detailed information about the
error messages.
If the error is conrmable, try xing it and press conrm
Shut the computer down and restart it.
If the error message is still on the screen, contact Valon
Lasers or local distributor’s service and declare the
num-ber of the error message.
11. Troubleshooting