46
In case the system seems to operate improperly, this sec-
tion will help you to locate and repair the defect. If the trou-
bleshooting instructions below do not correct the problem,
contact the manufacturer or an authorized service repre-
sentative for further assistance.
11.1. System does not turn on
• Ensure the power cable is plugged in.
• Check that the key switch is in correct position.
• Ensure the monitor is started.
• Ensure that the emergency switch is not pressed
11.2. Aiming beam not visible
• Check the system is on
• Ensure you have pressed the laser ready button. When
laser is enabled, the aiming beam should be visible.
• If the aiming beam is not visible turn the aiming beam to
maximum power.
• Ensure the spot size is correctly selected.
• Restart the system by shutting down the PC and starting
it again.
• Check the optical ber is connected to SLA. If not, call
an authorized Valon service representative
• If problems still occur, contact an authorized Valon ser-
vice representative
11.3. Aiming beam not in focus
• Check the focusing stick is in focus. If not, adjust the
binocular lenses to meet your diopters. Set the left eye
rst and then the right eye.
• Check whether the slit is correctly in focus on the fo-
cusing stick. If the slit is in order contact Valon Lasers
or local distributor’s service for help. For advanced slit
lamp service contact your local Haag-Streit or CSO rep-
resentative.
11. Troubleshooting
Reference
See also slit
lamp’s opera-
tion manual for
help