Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
196 / 214
COPYRIGHT© 2020 VATECH.
Command Remark
[SMP_ECHK_0002] At the end of abnormal shooting by releasing the irradiation
switch, Messages sent before [SMP_STOP_0000]-Purpose
of identifying the cause of the outage
[SMP_CPOF]
This command is to turn off the power of the sensor and
notify it when there is no command input for 30 minutes.
[SPM_STIT_0000] Xxxx captures in CT stitching mode
PANO mode sensor reference position setting command
[SPM_PMCT_0000]
[SPM_CMCT_0000] CT mode sensor reference position setting command
[SPM_CSMV_0000] The command for setting collimator drive amount between
the inner circle and outer circle during CT mode capture
[SPM_CCMV_0000] The command for setting collimator drive amount between
the inner circle and outer circle during CT mode capture
[SPM_PSCR_0000] PANO mode sensor, collimator drive frequency ration setting
[SPM_CSCR_0000] CEPH mode sensor, collimator drive frequency ration setting
[SPM_CPIO_0000] PANO mode collimator’s capture initial position offset setting
[SPM_CHP?] Command to check chin rest position
[SPM_XRES] X-ray remote exposure start
[SPM_XREE] X-ray remote exposure end
[SPM_XRET] X-ray remote exposure time set (ex) 0500 = 5 sec
[SPM_XREH] X-ray remote exposure halt
[SPM_LONV] Vertical beam on
[SPM_LOFV]: Vertical beam off