Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
70 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
12. If it does not meet the standard, move the sensor with a command and adjust it.
When the ofs value increases, move the sensor to the ORG
direction (Right): [spm_CMCT_xxxx] value decrease
When the ofs value decreases, move the sensor in the opposite
direction to the ORG (Left): [spm_CMCT_xxxx] value increase
[spm_CMCT_xxxx] value increase
* To modify ofs value by 1, modify CMCT value by 2.
13. When the standard pass, enter the ofs value in the DETOFFSETYY parameter in the
reconstruction.vxm, reconstruction_Bead.vxm, and reconstruction_GEO.vxm files in the
following folder
Half beam
Path: C:\VCaptureSW\VCaptureSW\Acquisition\CBCT\WidePANO\REC
File: (1) reconstruction.vxm, (2) reconstruction_Bead.vxm, (3) reconstruction_GEO.vxm
Parameter: DETOFFSETYY = -xxx setting in (1) & (2) & (3) files
When the CMCT value is determined, enter the PMCT value as (CMCT value + 149).
ex. [SPM_CMCT_431] → [SPM_PMCT_580]
When moving the
sensor, be
sure to
check the sensor level.
Check the Ofs value and
move the sensor
according to the value.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals