Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
80 / 214
COPYRIGHT© 2020 VATECH.
10. If it is not aligned, loosen the fixed point of the chinrest drive and move the chinrest drive
module.
When rotating clockwise, the top pin moves to the left.
When rotating counterclockwise, the top pin moves to the right.
11. After alignment, recover the value by sending the original HFST value.
ex. If it is [SPM_HFST_3800], input [SPM_HFST_200].