Vitrek V4 – Electrical Safety Tester USER MANUAL
40
:FUNCtion:TEST:STATe
Function:
Set or query the test state ( 0:OFF 1:ON 2:CONTINUE)
Syntax:
:FUNCtion:TEST:STATe <NR1>
:FUNCtion:TEST:STATe?
<NR1> is in the range from 0 through 2.
Return:
<NR1>
Examples:
:FUNCtion:TEST:STATe 1 - start test.
:FUNCtion:TEST:STATe? - returns “1” if the test state is 1.
:MEASure
Function:
Return measurement value
(NOTE: if no test has been run, the value returned by this function will be all zeroes)
Syntax:
:MEASure <X>?
<X> is in the range from 0 through 16 for STEP.
You must enter a step number. Do not use a space between the command and the step number.
Return:
Status, Mode, V/I, Meas, Time
(2 bytes) (1 byte) (7 bytes) (8 bytes) (7 bytes)
Status: 00:Emtpy
10:Pass
20:Fail
21:Interlock
22:Low Voltage
23:Hi voltage
24:ARC
25:Measurement over limit maximum
26:Measurement below limit minimum
30:Stop
32:Hight Voltage
33:Heatsink is too hot
Mode: 0: Empty
1:ACW
2:DCW
3:IR
4:GB
5:CAC
6:CDC
7:HOLD