Vitrek V4 – Electrical Safety Tester USER MANUAL
41
V/I:
Test Function (5 bytes) Unit (2 bytes)
ACW KV
DCW KV
IR -V
GB -A
CAC KV
CDC KV
HOLD PAUSE
Meas:
Test Function (5 bytes) Unit (4 bytes)
ACW MA
DCW MA
IR M
GB m
CAC MA
CDC MA
HOLD
Time:
Time (1 byte) Value (5 bytes) Unit (1 byte)
Ramp 005.0 sec
Test 005.0 sec
Examples:
Use the command :MEAS3? after measurement
Display for PC 10,1,0.100kV,01.00mA, T005.0S
10=Pass
1=ACW
0.100kV=Measurement voltage 100V
01.00mA=Measurement current 1mA
T005.0S=Test time 5 sec
:UTILity:ACWStanding:FREQuency
:UTILity:ACContinuous:FREQuency
Function:
Set or query the ACW or CAC frequency (50 or 60Hz )
Syntax:
:UTILity:ACWStanding:FREQuency <NR>
:UTILity:ACWStanding:FREQuency?
<NR>can be 50 (50Hz) or 60 (60Hz)
Return:
50 | 60
Examples:
:UTILity:ACWStanding:FREQuency 50 - sets the ACW frequency to 50Hz
:UTILity:ACWStanding:FREQuency? - returns “50” if the ACW frequency setting is 50Hz.
(NOTE: The Test Mode must be set to either ACW or CAC before this setting can be changed)