Vitrek V4 – Electrical Safety Tester USER MANUAL
42
:UTILity:ARC:CURRent
Function:
Set or query the ARC current
Syntax:
:UTILity:ARC:CURRent <NR2>
:UTILity:ARC:CURRent?
<NR2> is in the integer range 1 to 40 mA
Return:
<NR2>
Examples:
:UTILity:ARC:CURRent 2 - sets the ARC current to 2.00mA.
:UTILity:ARC:CURRent? - returns “2.00” if the ARC current setting is 2.00mA.
:UTILity:ARC:MODe
Function:
Set or query the ARC test state
(0:disable, 1:Enable & Stop, 2:Enable & Continue)
Syntax:
:UTILity:ARC:MODe <NR1>
:UTILity:ARC:MODe?
<NR1> is in the range from 0 through 2.
Return:
<NR1>
Examples:
:UTILity:ARC:MODe 0 - sets the ARC test state to disable.
:UTILity:ARC:MODe? - returns “0” if the ARC mode setting is disable.
:UTILity:TEST:MODe
Function:
Set or query the test mode ( 0:From Step 1, 1: From the present step )
Syntax:
:UTILity:TEST:MODe <Boolean>
:UTILity:TEST:MODe?
<Boolean>can be 0 or 1
Return:
0 | 1
Examples:
:UTILity:TEST:MODe 0 - sets the test mode to “From Step 1”.
:UTILity:TEST:MODe? - returns “0” if the mode setting is From Step 1.