Vitrek V4 – Electrical Safety Tester USER MANUAL
43
:UTILity:IRTest:MODe
Function:
Set or query the IR test mode ( 0:Stop on Fail, 1:Stop on Pass, 2:TIMER )
Syntax:
:UTILity:IRTest:MODe <NR1>
:UTILity:IRTest:MODe?
<NR1> is in the range from 0 through 2.
Return:
<NR1>
Examples:
:UTILity:IRTest:MODe 0 - sets the IR test mode to “Stop on Fail”.
:UTILity:IRTest:MODe? - returns “0” if the IR test mode setting is Stop on Fail.
:UTILity:GBZero:TEST
Function:
Start the GB zero check.
Syntax:
:UTILity:GBZero:TEST
:UTILity:GBZero:TEST?
Return:
Examples:
:UTILity:GBZero:TEST - Start the GB zero check.
:UTILity:GBZero:STATe
Function:
The query will return “1” if the GB zero check is OK and “0” if the GB zero check is fail.
Syntax:
:UTILity:GBZero:STATe?
Return:
0|1
Examples:
:UTILity:GBZero:STATe? - returns “1” if the GB zero check is OK.
:ACContinuous:VOLTage
Function:
Set or query the CAC test voltage.
Syntax:
:ACContinuous:VOLTage <NR2>
:ACContinuous:VOLTage?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACContinuous:VOLTage 1.000 - sets the CAC test voltage to 1.000kV.
:ACContinuous:VOLTage? - returns “1.000” if the CAC test voltage setting is 1.000kV.