Vitrek V4 – Electrical Safety Tester USER MANUAL
45
:DCContinuous:VOLTage
Function:
Set or query the CDC test voltage.
Syntax:
:DCContinuous:VOLTage <NR2>
:DCContinuous:VOLTage?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinuous:VOLTage 1.000 - sets the CDC test voltage to 1.000kV.
:DCContinuous:VOLTage? - returns “1.000” if the CDC test voltage setting is 1.000kV.
:DCContinuous:CMINimum
Function:
Set or query the CDC test minimum current.
Syntax:
:DCContinuous:CMINimum <NR2>
:DCContinuous:CMINimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinuous:CMINimum 1.00 - sets the CDC test minimum current to 1.00mA.
:DCContinuous:CMINimum? - returns “1.00” if the CDC test minimum current setting is 1.00mA.
:DCContinuous:CMAXimum
Function:
Set or query the CDC test maximum current.
Syntax:
:DCContinuous:CMAXimum <NR2>
:DCContinuous:CMAXimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinuous:CMAXimum 1.00 - sets the CDC test maximum current to 1.00mA.
:DCContinuous:CMAXimum? - returns “1.00” if the CDC test maximum current setting is 1.00mA.