Vitrek V4 – Electrical Safety Tester USER MANUAL
46
:DCContinuous:RTIMe
Function:
Set or query the CDC ramp time.
Syntax:
:DCContinuous:RTIMe <NR2>
:DCContinuous:RTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinuous:RTIMe 5.0 - sets the CDC ramp time to 5 sec.
:DCContinuous:RTIMe? - returns “5” if the CDC ramp time setting is 5 sec.
:ACWStanding:VOLTage
Function:
Set or query the ACW test voltage.
Syntax:
:ACWStanding:VOLTage <NR2>
:ACWStanding:VOLTage?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
: ACWStanding:VOLTage 1.000 - sets the ACW test voltage to 1.000kV.
: ACWStanding:VOLTage? - returns “1.000” if the CAC test voltage setting is 1.000kV.
:ACWStanding:CMINimum
Function:
Set or query the ACW test minimum current.
Syntax:
:ACWStanding:CMINimum <NR2>
:ACWStanding:CMINimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACWStanding:CMINimum 1.00 - sets the ACW test minimum current to 1.00mA.
:ACWStanding:CMINimum? - returns “1.00” if the ACW test minimum current setting is 1.00mA.