Vitrek V4 – Electrical Safety Tester USER MANUAL
47
:ACWStanding:CMAXimum
Function:
Set or query the ACW test maximum current.
Syntax:
:ACWStanding:CMAXimum <NR2>
:ACWStanding:CMAXimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACWStanding:CMAXimum 1.00 - sets the ACW test maximum current to 1.00mA.
:ACWStanding:CMAXimum? - returns “1.00” if the ACW test maximum current setting is 1.00mA.
:ACWStanding:RTIMe
Function:
Set or query the ACW ramp time.
Syntax:
:ACWStanding:RTIMe <NR2>
:ACWStanding:RTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACWStanding:RTIMe 5.0 - sets the ACW ramp time to 5 sec.
:ACWStanding:RTIMe? - returns “5” if the ACW ramp time setting is 5 sec.
:ACWStanding:TTIMe
Function:
Set or query the CAW test time.
Syntax:
:ACWStanding:TTIMe <NR2>
:ACWStanding:TTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACWStanding:TTIMe 5.0 - sets the ACW test time to 5 sec.
:ACWStanding:TTIMe? - returns “5” if the ACW test time setting is 5 sec.