Vitrek V4 – Electrical Safety Tester USER MANUAL
48
:DCWStanding:VOLTage
Function:
Set or query the DCW test voltage.
Syntax:
:DCWStanding:VOLTage <NR2>
:DCWStanding:VOLTage
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
: DCWStanding:VOLTage 1.000 - sets the DCW test voltage to 1.000kV.
: DCWStanding:VOLTage? - returns “1.000” if the DCW test voltage setting is 1.000kV.
:DCWStanding:CMINimum
Function:
Set or query the DCW test minimum current.
Syntax:
:DCWStanding:CMINimum <NR2>
:DCWStanding:CMINimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:CMINimum 1.00 - sets the DCW test minimum current to 1.00mA.
:DCWStanding:CMINimum? - returns “1.00” if the DCW test minimum current setting is 1.00mA.
:DCWstanding:CMAXimum
Function:
Set or query the DCW test maximum current
Syntax:
:DCWStanding:CMAXimum <NR2>
:DCWStanding:CMAXimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:CMAXimum 1.00 - sets the DCW test maximum current to 1.00mA.
:DCWStanding:CMAXimum? - returns “1.00” if the DCW test maximum current setting is 1.00mA.