Vitrek V4 – Electrical Safety Tester USER MANUAL
49
:DCWStanding:RTIMe
Function:
Set or query the DCW ramp time.
Syntax:
:DCWStanding:RTIMe <NR2>
:DCWStanding:RTIMe
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:RTIMe 5.0 - sets the DCW ramp time to 5 sec.
:DCWStanding:RTIMe? - returns “5.0” if the DCW ramp time setting is 5 sec.
:DCWStanding:TTIMe
Function:
Set or query the DCW test time.
Syntax:
:DCWStanding:TTIMe <NR2>
:DCWStanding:TTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:TTIMe 5.0 - sets the DCW test time to 5 sec.
:DCWStanding:TTIMe? - returns “5.0” if the DCW test time setting is 5 sec.
:IRESistance:VOLTage
Function:
Set or query the IR test voltage.
Syntax:
:IRSistancce:VOLTage <NR1>
:IRSistancce:VOLTage?
<NR1> Please refer to the specification
Return:
<NR1> is 50, 100, 500, or 1000
Examples:
:IRSistancce:VOLTage 100 - sets the IR test voltage to 100V.
:IRSistancce:VOLTage? - returns “100” if the IR test voltage setting is 100V.