Stimulation Tab
Detection Tab
2
Tachycardia Detection
You may enable or disable Detection. If Detection is Disabled, then the model
1000 generators use only Normal and Magnet stimulation. If Detection is
Enabled, then parameters for AutoStim will become available, in addition to
Normal and Magnet parameters.
When you enable Detection for the first time, the software will prompt you to
set the Heartbeat Detection setting and AutoStim Threshold. These settings
work together to ensure the generator is accurately detecting the patient’s
heartbeats, and set the criterion for AutoStim delivery based on changes in
heart rate, respectively. Once Detection is enabled, you can adjust the settings
from the Detection tab as needed.