34
To exit Scheduled Programming, select EXIT GUIDED.
NOTE:
If the patient resides in a di erent time zone or plans to travel, program
the schedule based on the patient’s local time zone to ensure therapy
changes occur at the intended times.
NOTE:
Protocols that utilize 0.25mA increases such as the Standard Protocol
require a minimum of 14 days between scheduled automatic titration
steps. Custom protocols that utilize 0.125mA increases require a 7-day
minimum interval between scheduled titration steps.
Scheduled Apply