EasyManua.ls Logo

VWR Puranity TU3+ - Chapter 5 Technical Specification

Default Icon
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VWR Puranity TU+ | 13
Technical specification
NOTE!
Check at regular intervals the quality of your feedwater.
Feedwater requirements
Source Tap water
Silt Density Index (SDI) Bacteria should be < 0.01
Conductivity (reference temperature +25 °C) 1200 μS/cm (0.0008 M
·cm)
Free chlorine max. 0.1 ppm
Manganese content max. 0.05 ppm
Iron content max. 0.05 ppm
pH range 4 - 11
Temperature +2°C - +35°C
Pressure 1 - 6 bar /14 - 87 PSI (at a pressure > 6 bar / 87 PSI a pressure
reducer must be installed upstream of the system.
Product water quality ASTM Type I (on dispensing valve)
Standard UV UV/UF
Conductivity
(reference temperature +25 °C) μS/cm 0.055 0.055 0.055
Resistance
(reference temperature +25°C) M
·cm 18.2 18.2 18.2
TOC value ppb 5-10 1-5 1-5
RNase
DNase
ng/ml
pg/ul
- -
- -
- -
- -
< 0.003
< 0.4
Bacteria CFU/ml < 0.01 < 0.01 < 0.01
Bacterial endotoxins EU/ml - - - - < 0.001
Particles μm/ml < 0.2 < 0.2 < 0.2
Performance l/min 1.0 1.0 0.6

Table of Contents