| The installation area
20
|
Puranity TU+ VWR
Chapter 6
An atmospherically floor drain with an outside diameter of 63 mm or 2.48 inch (DN 50
tube) shall be provided.
Unobstructed draining of the rinsing water to the drain must be ensured.
When no floor drain is available, install a water watcher to protect against water damage
(available only for Europe).
A check valve is recommended in the feedwater line to prevent back flow of feedwater
from the water system.
An electric socket with protective connection must be available for connection of the
system to the voltage supply. (see “Electrical connections” on page 16).
Ample working space must be provided around the system for easy and pleasant
replacement of wear and replacement parts and for ease of operation (see “Accessibility
to Puranity TU+ systems and pure water tank” on page 17).
Easy access for operation and control of the system.