VWR Puranity TU+ | 51
Maintenance |
Chapter 13
3 Press on the two quick connectors at
the inlet and outlet of the ultrapure
cartridge and remove the used
cartridge from the system.
4 We recommend performing a cleaning
when an existing ultrapure cartridge is
replaced.
NOTE!
Refer to section “Cleaning” on page 53 for an explanation of
how to perform cleaning.
5 Locate the new ultrapure cartridge and
insert the cartridge into the system.
6 Insert the quick connectors into the
connecting points on the ultrapure
cartridge. When you hear an audible
click you can be sure that the quick
connectors have been inserted
correctly.
NOTE!
The quick-connectors are attached
to the system in such a manner so it
is to prevent any confusion
(switching).
7 Put the cover for the ultrapure
cartridge back in place and switch the
system on (connect the main power
supply).
8 NOTE!
Discharge the first 5 liters.
Step Action Figure