(05)
Alkalinity-M
5 - 200 mg/l (CaCO₃)
Tablet
Internal Name: 05-Alkalinit-M-tab
10ml
Alkalinity-M Photometer (TbsPTA)
Measurement procedure:
1 Fill 10 ml test water into a clean 24 mm cuvette.
2 Screw the lid back on the cuvette.
3 Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4 Start ZERO measurement.
5 Remove the cuvette again.
6 Unscrew the lid from the cuvette.
7 Add 1 Alkalinity-M Photometer tablet(s) to the test water in the cuvette.
8 Crush the tablet with a clean stirring rod.
9 Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
10 Screw the lid back on the cuvette.
11 Gently swirl the cuvette to mix the liquid well.
12 Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
13 Tap TEST to start a 00:25 minute(s) countdown.
14 After the lapse of a 00:25 minute(s) countdown the measured value is displayed.
Notes:
The test result can be converted into the following unit(s): mg/l HCO₃⁻, °dH, °eH, °fH, mmol
(KS4.3), mval.
T3/204 T3/204

Other manuals for Water I.D. Primelab 2.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Water I.D. Primelab 2.0 and is the answer not in the manual?

Water I.D. Primelab 2.0 Specifications

General IconGeneral
BrandWater I.D.
ModelPrimelab 2.0
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals