EasyManuals Logo

Water I.D. Primelab 2.0 User Manual

Water I.D. Primelab 2.0
301 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
(63) Bromine
(in absence of chlorine)
0.00 - 9.00 mg/l (fBr₂)
Liquid
+ Powder
Internal Name: 63-Bromine-liq
10ml
30ml PL DPD 1 A (PL30DPD1A)
30ml PL DPD 1 B (PL30DPD1B)
PL DPD Nitrite Powder (PLpow20DPDNitr)
Measurement procedure:
1 Fill 10 ml test water into a clean 24 mm cuvette.
2 Screw the lid back on the cuvette.
3 Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4 Start ZERO measurement.
5 Remove the cuvette again.
6 Unscrew the lid from the cuvette.
7 Empty the cuvette to a few drops.
8 Add 3 drop(s) of PL DPD 1 A into the cuvette.
9 Add 3 drop(s) of PL DPD 1 B into the cuvette.
10 Fill the cuvette to 10 ml with the sample water.
11 Screw the lid back on the cuvette.
12 Gently swirl the cuvette to mix the liquid well.
13 Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
14 Tap TEST to perform the measurement.
15 The measured value for tBr₂ (total bromine) is immediately displayed.
16 If you require determination of ´combined´ and/or ´free´ bromine, please proceed with the
following steps.
17 Remove the cuvette again.
18 Empty the cuvette.
19 Clean the cuvette.
20 Fill 10 ml sample water into a second clean 24 mm cuvette.
21 Add 1 x 0.05mL PL DPD Nitrite scoop(s) powder to the sample water in the cuvette.
22 Screw the lid back on the cuvette.
23 Gently swirl the cuvette to mix the liquid well.
24 Add 3 drops of PL DPD 1 A into a second clean 24 mm cuvette.
25 Add 3 drop(s) of PL DPD 1 B into the cuvette.
26 Fill the cuvette with 10 ml of the treated sample water from the first cuvette.
T15/204 T15/204

Other manuals for Water I.D. Primelab 2.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Water I.D. Primelab 2.0 and is the answer not in the manual?

Water I.D. Primelab 2.0 Specifications

General IconGeneral
BrandWater I.D.
ModelPrimelab 2.0
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals