Table B–5: Mobile phase absorbance measured against air or water (continued)
Absorbance (AU) at specified wavelength (nm)
200 205 210 215 220 230 240 250 260 280
Waters PIC reagents
PIC A, 1 vial/L 0.67 0.29 0.13 0.05 0.03 0.02 0.02 0.02 0.02 <0.01
PIC B6, 1 vial/L 2.46 2.50 2.42 2.25 1.83 0.63 0.07 <0.01
PIC B6, low UV, 1 vial/L 0.01 <0.01
PIC D4, 1 vial/L 0.03 0.03 0.03 0.03 0.02 0.02 0.02 0.02 0.02 0.01
BRI J 35, 1% 0.06 0.03 0.02 0.02 0.02 0.01 <0.01
3-[(3-cholamidopropyl)-
dimethylammonio]-1-propanesulfonate)
(CHAPS), 0.1%
2.40 2.32 1.48 0.80 0.40 0.08 0.04 0.02 0.02 0.01
Sodiumdodecyl sulfate (SDS), 0.1% 0.02 0.01 <0.01
4-octylphenol polyethoxylate (Triton X-100),
0.1%
2.48 2.50 2.43 2.42 2.37 2.37 0.50 0.25 0.67 1.42
Polyoxyethylene sorbitan monolaurate
(Tween 20), 0.1%
0.21 0.14 0.11 0.10 0.09 0.06 0.05 0.04 0.04 0.03
June 12, 2015, 715004747 Rev. A