Directions for Use Chapter 2 Setup 25
MAP Measurement Enable and Disable
Default setting: MAP ENABLED for neonate; MAP DISABLED for adult and pediatric.
1. Depress for 3 seconds. SET HOUR XX appears in the message window.
2. Press repeatedly until MAP ENABLED or MAP DISABLED appears in the display
window.
3. Press or to enable or disable MAP measurement.
4. To select the displayed state and return to normal operation, either do nothing for 10
seconds or press any button other than or . If you press a function button (such
as ), the monitor returns to normal operation with that function ( ) activated.
For information about MAP measurements, see “Patient Monitoring” on page 31.
Note
If you change the MAP enabled/disabled setting, refer to “How Changing the
Patient Type Affects MAP Defaults” on page 38.
SYS
DIA
SpO2 %
/min
ºF
ºF
mmHg
MAP DISABLED
SYS
DIA
SpO2 %
/min
ºF
ºF
mmHg
MAP ENABLED