82 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series
Factory Default Settings
Function Value
Blood pressure measurement units mmHg
Patient type Adult
Automatic NIBP Interval 15 minutes
MAP (Adult and Pediatric) Disabled
MAP (Neonate) Enabled
NIBP Adult High Systolic Alarm 220 mmHg
NIBP Adult Low Systolic Alarm 75 mmHg
NIBP Adult High Diastolic Alarm 110 mmHg
NIBP Adult Low Diastolic Alarm 35 mmHg
NIBP Adult High MAP Alarm 120 mmHg
NIBP Adult Low MAP Alarm 50 mmHg
NIBP Pediatric High Systolic Alarm 145 mmHg
NIBP Pediatric Low Systolic Alarm 75 mmHg
NIBP Pediatric High Diastolic Alarm 100 mmHg
NIBP Pediatric Low Diastolic Alarm 35 mmHg
NIBP Pediatric High MAP Alarm 110 mmHg
NIBP Pediatric Low MAP Alarm 50 mmHg
NIBP Neonatal High Systolic Alarm 120 mmHg
NIBP Neonatal Low Systolic Alarm 50 mmHg
NIBP Neonatal High Diastolic Alarm 70 mmHg
NIBP Neonatal Low Diastolic Alarm 30 mmHg
NIBP Neonatal High MAP Alarm 80 mmHg
NIBP Neonatal Low MAP Alarm 35 mmHg
SpO
2
Adult High Alarm ---
SpO
2
Adult Low Alarm 85%
SpO
2
Pediatric High Alarm ---
SpO
2
Pediatric Low Alarm 85%
SpO
2
Neonatal High Alarm 95%
SpO
2
Neonatal Low Alarm 85%
Temperature Scale °F (Fahrenheit predictive)
Pulse Rate Adult High Alarm 120 beats per minute
Pulse Rate Adult Low Alarm 50 beats per minute
Pulse Rate Pediatric High Alarm 150 beats per minute
Pulse Rate Pediatric Low Alarm 50 beats per minute
Pulse Rate Neonatal High Alarm 200 beats per minute
Pulse Rate Neonatal Low Alarm 100 beats per minute
Pulse Tone Volume 03
Print Control Batch
Time-of-day Display 24-hour