Directions for Use Chapter 4 Alarms and Alerts 55
Nonrecoverable Alerts
When the monitor detects a nonrecoverable equipment problem, it does the following
Displays an error code in the SYS window, and shuts off the display to all other
windows
Stops patient monitoring
Stops the pump and opens the air valve
Activates the Nurse Call relay (if connected)
Produces an audible tone
Shuts down as soon as is pressed or one minute has elapsed
Response
For both recoverable and nonrecoverable equipment alerts, respond as follows:
1. Press to immediately silence the alert tone.
2. Determine what caused the alert and correct the problem.