Directions for Use Chapter 3 Patient Monitoring 47
Axillary Predictive
When used correctly, the monitor produces an accurate axillary temperature
measurement in less than 15 seconds for adults and in less than 13 seconds for pediatric
patients.
Use Axillary Pediatric (AP) measurements for patients up to 17 years old.
Use Axillary Adult (AA) measurements for patients 18 years old and older.
1. Remove the temperature probe from the probe holder.
The temperature probe runs a self-test, displaying 188.8 for a few seconds. When it is
ready for use, the temperature window clears, and then OrL appears in the
temperature window.
2. Press or to change the display to AP or AA.
3. Load a new probe cover by inserting the probe into a probe cover and pressing the
probe handle down firmly. The probe handle moves slightly to engage the probe
cover.
4. Lift the patient’s arm to fully expose the axilla.
Note
For axillary temperatures, use only the oral probe (blue ejection button) and the
blue probe well.
Caution Use only Welch Allyn probe covers. The use of any other probe cover, or
failing to use a probe cover, can produce measurement errors or inaccuracies.
Note
Be sure that nothing touches the probe tip before you place it in the axillary
measurement site.
Note
Do not allow the probe tip to make contact with the patient until the probe is
placed in the measurement site. Any such contact can cause an inaccurate
reading.