Section 2 - Service
SERVICE MANUAL 6200-43E REV. A WELCH ALLYN ATLAS MONITOR 13
2.3.7Set battery charging voltage:
Specification :No Load: 6.85VDC
Need: DVMM
2.3.7.1Remove battery from unit and unplug.
2.3.7.2Use DVMM to check across right pin (+)
and left pin (-) when viewed looking into
the battery compartment.
2.3.7.3Adjust potentiometer R338 to obtain
6.85VDC. (It is located at the right of the
battery jack.
Turn it counterclockwise to increase the
charging voltage.
2.3.7 Set Battery Charge
Voltage