2 WELCH ALLYN ATLASMONITOR
Section 4 - Disassembly and Repair, Domestic
SERVICE MANUAL 6200-43E REV. A
4.1.1 Open the paper door and remove the
paper roll. Close paper door.
.
4.1.2 Remove 4 Torx Machine screws with
Torx - 10 screw driver.
4.1.3 Open the rear housing and
disconnect 2 ground wires from
ground lug as well as printer cable..
4.1 MODEL 200 Dissassembly.
Unplug Monitor from
Mains before disassem-
bly! EXERCISE EXTREME
CAUTION WHEN SERVICING
THE ATLAS MONITOR! THE
CRT ASSEMBLY AND CRT DE-
FLECTION BOARD OPERATE
ON APPROXIMATELY 8,000
VOLTS.
ESD: Circuit boards are sensitive
to static electricity. Use wrist strap,
ESD mats and ESD storage bags.
Place Monitor on a foam pad to
protect the front of the unit and the
pressure fitting on the front of the
unit.
Caution: Do not pull display cable.
Model 200