EasyManua.ls Logo

Welch Allyn Atlas - 3.2 Chief Complaint, Cause and Corrective Action

Welch Allyn Atlas
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Complaint
Cause
Corrective Action
Power
Will not power up in AC
No wall power
Wrong wall power voltage/frequency
Fuse in power supply
Fuse in neutral wire
Power supply failure
LEDs
Random LSD segments unlit
Failed LED
LED 1
Subsystem
LEDs
unlit
Subsystem problem Check subsystem LED 1
LED intermittent, dim, flickering
Failed LED
Replace Display board
Buttons
Button not functional
Failed switch
Button 1
Failed subsystem
Check subsystem
Button sticking under front bezel
Possible loose display board
Tight display board mounting screws.
Button intermittent or difficult to
Failed switch
Button 1
make contact
sounds
No sound at all
Failed speaker or disconnected
Failed main board
Software corruption
Reload software
sound 1
Battery
Insufficient
life
Printer
Feed problems
Waveform print quality
Text print quality
Not functional
Feeding but not printing
Failing battery
Battery 1
Paper inserted incorrectly
Door not latched
Failed printer
Software adjustment needed
Software adjustment needed
Failed printer
Printer cable
Paper inserted backwards
Wrong kind of paper
Failed printer
Printer 1
Printer 1
Printer 1
Printer 1
Printer 1
CRT
CRT is blank
CRT cables
Failed CRT
Failed Deflection board
Failed Main Board
CRT 1
Reinstall software
Section 3 -Troubleshooting
3.2 Chief Complaint, Cause and Corrective Action
12 WELCH ALLYN ATLAS MONITOR
Service Manual
6200-43E
Rev. A

Other manuals for Welch Allyn Atlas

Related product manuals